linearity, LC detectors, detectors, detection, linear range, dynamic range, range, dynamic, linear, American Society for Testing and Materials, ASTM, performance, HPLC, high performance liquid chromatography, concentration, off-line test, static, static test, on-line test, dynamic test, non-linear detector, UV, UV detector, output, slope, line, equation, sample, chromatography, chromatographers, relationship, response, chemistry, analytical chemistry, analytical, university of kentucky, college of pharmacy 

Linearity In LC Detectors

A detector operates by registering an output in response to sample detection. Chromatographers expect a linear relationship between response of the detector and concentration of the sample, and calibration techniques are designed to promote this relationship. However, not all detectors are linear and effects from other components of the HPLC system may cause the detector to deviate from a linear response between response and concentration. The following section discusses how a detector's linearity can be measured and evaluated, and it provides illustrations of these techniques. Much of this section follows guidelines and terminology adopted by the American Society for Testing and Materials (ASTM).Two important terms when describing detector performance are linear range and dynamic range. Linear range is based upon the following equation:For a linear detector the equation beomces R = SC since Ro should equal zero. A plot of response versus concentration would result in a line with slope = S, where the value of S = R/C. Thus, a linear detector is one where the sensitivity is constant at all sample concentrations. Subsequently, linear range refers to concentrations where S is constant within a certain tolerable range, usually +/- 5% (Dorschel, 1989). ASTM refers to dynamic range as "that range of concentrations of the test substance, over which a change in concentration produces a change in detector signal." The lower limit of the dynamic range is the concentration where the detector signal is equal to twice the detector's noise level (random variations in detector signal). The upper limit is the concentration where the slope of the detector response curve approaches zero. If the curve does not have a plateau, then the highest measured concentration is understood to be the upper limit.When measuring and evaluating linearity in detectors, off-line (static) test or on-line (dynamic) test is used. The static test is where the detector is disconnected from the HPLC system and the sample concentrations are injected directly into the detector cell via a syringe, pump, etc., where no constant flow through the detector is seen. It is the procedure to use for nondestructive detectors and it allows the user to know the exact concentration of sample reaching the detector.On the other hand, the dynamic test is performed using the entire HPLC system. This is more suitable for destructive detectors, such as electrochemical, and it is easier to use. It also allows testing the detector to the lowest of its output response levels since any nonsample components or gases in the flow solvent can be resolved or separated from the desired sample.The process for measuring linear or dynamic ranges for nondestructive detectors involves using solutions with known dilutions of the sample compound. The sample compound must also have known detection properties, such as absorbance or refractive index. The solutions are injected into the detector, the detector response is recorded, and the detector is washed with pure solvent in between each test dilution. It should be noted that certain factors may change results of this test, such as choice of sample or detector operating conditions. Factors such as these are beyond the scope of this section and will not be discussed.Using ASTM guidelines, Dorshel et al. evaluated the linearity of an anonymous UV detector using acenaphthene as the test sample and employing a static system (Dorschel, 1989). The following response vs. concentration plot was obtained:(Dorschel, 1989)Graph 1 would seem to imply a linear plot since all but the final data point falls on the line beginning at the origin and extrapolated out. However, subsequent results seemed to prove that the chromatographer needs to view the experimental results more closely before a conclusion of linearity for this detector can be decided.Graph 1 shows that the dynamic range upper limit is 0.25 g/L (the highest measured concentration since no plateau) and Graph 2 (logarithmic axes) shows the lower limit to be 1.75 X 10-6 g/L. Thus, the dynamic range, expressed as a ratio of the upper and lower limits, is 1.43 X 105 .The linear range was found using ASTM guidelines since all but one data point was included on the extrapolated line. First, the data points are connected using a smooth curve. Next, a best-fit line is drawn from the origin through the data points. Finally, another line with 95% slope of the best-fit line is drawn from the origin and extrapolated out. The concentration for the upper limit of the linear range is the point where the smooth curve intersects the 95% slope line:  (Dorschel, 1989)The lower limit is the same as with the dynamic range. Thus, the linear range for the ASTM procedure is 1.36 X 105.Dorshel et al. note that the ASTM method for linear range assumes that the detector is linear and the method may tend to smooth data which comes from a non-linear detector. Therefore, they calculated the linear range using the definition stated earlier, where the sensitivity values must fall within a certain tolerable region (+/- 5%). For this UV detector, the response should adhere to the Beer-Lambert Law: Since absorbance is equal to detector response, and the product of the extinction coefficient and optical pathlength, b, is equal to sensitivity, then R/C = A/C = Extin.Coeff. x b = S. So a R/C vs. Log C plot can be made, where the +/- 5% tolerance region (dotted lines) is chosen arbitrarily (in this example, 0.031g/L was chosen):  (Dorschel, 1989)The linear range in this technique (the concentrations which fall within the dotted lines) is 9.52. The linearity plot indicates that the detector is only linear over a certain, narrow range of concentrations. The significance of this result is not to omit use of this detector at certain concentrations, but instead to allow the chromatographer to recognize nonlinearity. Dorshel et al. insist that "it is not the existence of nonlinearity but rather the analyst's failure to anticipate and adjust for nonlinear behavior that can lead to poor quantification" (Dorschel, 1989)

.Dorschel, C.A.; Ekmanis, J.L.; Oberholtzer, J.E.; Warren, F.V. Jr.; Bidlingmeyer, B.A. Analytical Chemistry, 1989, Vol. 61, pp. 952.

Dorschel, p. 951A-968A.

Dorschel, p. 953A.

Dorschel, p. 954A.

Dorschel, p. 954A.

Dorschel, p. 956A.

Return to Detectors

Return to Theory of Operation Menu

Return to Users Guide Page

Exit to the University of Kentucky College of Pharmacy

linearity, LC detectors, detectors, detection, linear range, dynamic range, range, dynamic, linear, American Society for Testing and Materials, ASTM, performance, HPLC, high performance liquid chromatography, concentration, off-line test, static, static test, on-line test, dynamic test, non-linear detector, UV, UV detector, output, slope, line, equation, sample, chromatography, chromatographers, relationship, response, chemistry, analytical chemistry, analytical, university of kentucky, college of pharmacy