
Analysis of Energy Expenditure
(The slides used in this presentation can be viewed by clicking on the hyperlinks.)
Lipids in biological samples, including grains and meats, have been determined using
near-infrared methods for a number of years. The near-infrared spectrum of oleic acid shows the
distinctive spectral regions that correspond to lipids. The peak structures around 1200 nm,
1700-1800 nm, 2100-2200 nm, and 2300-2400 nm, and the lack of peaks elsewhere in the
spectrum are powerful indicators of the presence of lipids in a sample. The shifting of peaks in
these regions to lower energies (longer wavelengths) generally indicates cis-to-trans
isomerization or saturation of the lipid carbon-carbon double bond. Fatty acids like oleic acid
have broad regions of the spectrum with relatively low absorbance, enhancing tissue penetration
of near-IR light with these wavelengths.
To compare the performance of near-IR spectrometric imaging with a camera to direct
measurements made with a conventional diffuse reflectance spectrometer, an experiment was
performed with samples containing known amounts of different lipid mixtures. This picture in
visible-light of four different rabbit chows and two optical reflectance standards (one black and
one white, in the center of the image) was also imaged in the near-infrared using conventional
tungsten blackbody light bulbs. The four different chow formulations were high cholesterol
(1%), low fat, high saturated fat, and control. The purpose of the experiment was to determine
whether camera imaging at six wavelengths was as effective at differentiating among the chows
as scanning individual pellets in a diffuse reflectance spectrometer.
Multidimensional distances in standard deviations (SDs) were measured between samples of
chow with the near-IR camera at six different near-IR wavelengths. In the table, distances less
than 3 SDs denote a match between the samples, while distances greater than 3 SDs denote
samples with statistically different spectra. The near-IR camera was able to identify perfectly all
of the calibration and validation chow samples.
Similar distances in SDs between the chow samples were measured using a conventional near-IR
diffuse reflectance spectrometer. The conventional spectrometer also achieved perfect
identification of the different chow samples, in many cases with larger distances in SDs than the
camera. This increased discrimination ability is not unusual because the conventional
spectrometer is able to scan 701 wavelengths on a single chow pellet, with more even and
controlled illumination of all pellet surfaces. However, in the cross validation analysis one chow
pellet failed cross validation in the high cholesterol chow group. The failure to cross validate
was probably due as much to the small size of the calibration and validation pellet groups (nc=5,
nv=5) as to the increased sensitivity of the conventional spectrometer.
Determination of Heat from Temperature Measurements. Given heat capacity and mass, heat
content can be determined from temperature measurements. Surface temperature can be
determined from infrared photon (blackbody) emission. Infrared photon emission can be
detected using infrared cameras based on InSb and PtSi CCD technology. In this manner, the
energy output of an subject like a rat can be determined.
Two models are being used by our research group to predict core temperatures in metabolism
studies from the surface temperatures of different body parts. The models are based on
empirical data measurements or on computer simulations aided by empirical measurements.
Empirical Model
The regression of dose of a prototype drug against the temperature of various body zones, like
the tail, dorsal region, and core temperature, yields an equation that can be used to predict total
energy output. From similar dose measurements on new compounds, and images of skin
temperatures from infrared (IR) photon emission, core temperatures can be predicted accurately.
This type of empirical modeling has been demonstrated in the metabolism literature in rats and
in human subjects.
Simulation Model
The simulation model is a semi-empirical model that uses the laws of heat transfer, bulk
temperature and metabolic measurements, and a three-dimensional model of the body to
calculate spatially resolved temperatures from the input data (including image data).
This IR emission image of an alert, freely moving rat shows areas with the highest surface
temperatures in shades of red, and proceeds through the spectrum to portray decreasing
temperatures in shades of yellow, green, blue, and violet. Areas with temperatures off the
calibration scale on the high end are colored in white, while areas with temperatures off the
calibration scale on the low end are colored in black.
Tissue temperature control is established by a balance between energy production and energy
dissipation.
Heat conduction between materials is governed by an equation based on the temperature,
thermal conductivity, area, and distance of conduction.
The variables and constants in equations governing heat transfer are given in this Table of
Constants.
Heat flow in joules can also be calculated by an equation based on the specific heat, mass, and
temperature change.
Heat convection can be modeled as heat conduction into a moving material followed by heat
conduction out of the moving material and into another substance. This is similar in concept to
the three-layer insulation problem that often appears in textbooks.
Heat radiation is governed by an equation known as Stefan's Law, which gives power output in
J/sec given the surface emissivity, area, temperature, and temperature of the surrounding
environment. Most of the heat lost by an subject like a rat can be calculated from heat
conduction, convection, and radiation, However, some heat is carried off by vaporization of
water. This amount of energy loss can be calculated using the latent heat of vaporization and
mass of water lost.
Other measurements can be used in the total energy balance calculation to estimate of energy
expenditure in chemical reactions. Oxygen consumption (by atmospheric Raman spectrometry),
oxy- and deoxyhemoglobin (by short-wavelength near-IR spectrometry of electronic transitions),
carbon dioxide and water (by infrared spectrometry) can also be made on the subject or the
surrounding atmosphere to assist in the determination of overall energy balance. Such
measurements typically require that the subject be housed in a metabolic calorimetry chamber.
The metabolic calorimetry chamber is a hybrid metabolic cage and Dewar, constructed from
reflective aluminum and polystyrene foam panels. The chamber has multiple ports that can be
instrumented and used for near-IR and visible-light cameras, air inlets and outlets, urine
collection, food and water, temperature sensors, lighting, etc. The camera on the left of this
image is a Mitsubishi PtSi CCD IR camera, while the camera on the right is a Cincinnati
Electronics InSb camera converted to near-IR use. The top ports in the chamber are open, while
the ports on the sides are still screwed on.
The semi-empirical computer model for heat transfer partitions the total energy information
provided by temperature sensors, oxygen consumption, CO2 and water measurements into body
zones using the guidance of temperature images provided by IR photon emission.
Conjugate-gradient or other optimization of the semi-empirical model parameters to the
observed temperature data allow the model parameters to be adjusted until an adequate estimate
of energy expenditure is obtained.
The objective of the semi-empirical modeling is to allow reasonable estimates of spatially
resolved energy expenditure in freely moving subjects to be made. The accuracy of the
semi-empirical modeling is dependent upon the coarseness of the three-dimensional
grid(determined by the volume element, or voxel, size) used in the simulation. Finer gridspermit
more accurate simulation of energy transfer, in part because the dx terms in heat conduction
imply the use of an infinitesimally small volume element.
In a 3-D cubic grid, each cell has 26 nearest-neighbors. Changes in the temperature value in a
given cell propagate through the other cells according to the conduction, convection, radiation,
and vaporization rules established for the individual cells. The propagation of changes through
the grid with each time tick follows a wave-front equation and integrated wave-front equation
that determines how many cells are affected at each time "tick."
The semi-empirical computer model indexes the variables in each cell with values that indicate
how the cells are permitted to transfer energy by conduction, convection, radiation, and in some
cases, vaporization.
Because computer memory is contiguous and linear, functions are derived to map the locations
of the 26 nearest-neighbors in 3-D to addresses in a 1-D array, creating a multi-linked list data
structure. The pointers in the links to the temperature values in each cell are also used as
pointers to the corresponding conduction, convection, radiation, and vaporization parameters
used by the semi-empirical model.
Assessment of the semi-empirical model fit to the observed temperature-image data occurs
through calculation of the difference image between the model and observed data, and
minimization of the sum of the squared residuals. The parameters in the are adjusted
semi-empirical model are adjusted to maximize the goodness-of-fit using an optimization
method such as conjugate-gradient or simplex optimization. Monte Carlo ensemble forecasting
is used to gauge the robustness of the solution by randomly varying the initial parameters slightly
30 times and running the simulation repeatedly to see whether it always converges to the same
solution. (If 30 trials all reach the same solution, one can be much more confident of the results
than one could be if 30 trials converged on 20 or 30 different solutions.)
Return to ASRG Home Page