Drifting baseline: Causes and Solutions


                         Possible Cause                                                                                          Solution
 
 
Negative direction (gradient elution) - absorbance of mobile-phase A Use non-UV absorbing mobile phase solvents; use HPLC grade mobile phase solvents; add UV absorbing compound to mobile phase B. 
Positive direction (gradient elution) - absorbance of mobile phase B Use higher UV absorbance detector wavelength; use non-UV absorbing mobile phase solvents; use HPLC grade mobile phase solvents; add UV absorbing compound to modile phase A. 
Positive direction - contamination buildup and elution  Flush column with strong solvent; clean up sample; use HPLC grade solvents
Wavy or undulating - temperature changes in room Monitor and control changes in room temperature; insulate column or use column oven; cover refractive index detector and keep it out of air currents.