Noise: Causes and Solutions

                        Possible Cause                                                                                         Solution
 
 
Continous - detector lamp problem or dirty flow cell  Replace UV lamp( each should last 2000 h); clean and flush flow cell
Gradient or isocratic proportioning - lack of solvent mixing Use proper mixing device; check proportioning precision by spiking one solvent with UV absorbing compound and mointor UV absorbance detector output
Gradient or isocratic proportioning - malfunctioning proportioning valves Clean or replace proportioning precision valves; partially remix solvents
Occasional sharp spikes - external electrical interference Use voltage stabilizer for LC system; use independent electrical circuit
Periodic - pump pulses Service or replace pulse damper; purge air from pump; clean or replace check valves.
Random - contamination buildup Flush column with strong solvent; clean up sample; use HPLC grade solvent 
Spikes - bubble in detector  Degas mobile phase; use back pressure restrictor at detector outlet
Spikes - column temperature higher than boiling point of solvent Use lower column temperature