| Chelating solutes - trace metals in
base silica |
Use high purity silica-based column
with low trace-metal content; add EDTA or chelating compound to mobile phase; use
polymeric column |
| Silica-based column - degradation at
high pH |
Use polymeric, sterically protected,
or high-coverage reversed-phase column; install silica gel saturatorcolumn between pump
and injector |
| Silica-based column - degradation at
high temperature |
Reduce temperature to less than 50 C |
| Silica-based column - silanol
interactions |
Decrease mobile-phase pH to suppress silanol
ionization; increase buffer concentration; derivatize solute to change polar interactions |
| Unswept dead volume |
Minimize number of connections; ensure injector
rotor seal is tight; ensure all compression fittings arecorrectly seated |
| Void formation at head of column |
Replace column, or, if possible, open top
endfitting and clean and fill in void with glass beads or samecolumn packing;
rotate injection valve quickly; use injection valve with pressure bypass; avoid pressure
shock |