High back pressure: Causes and Solutions

                       Possible Cause                                                                                        Solution
 
 
Column blocked wth irreversibly adorbed sample Improve sample cleanup; use guard column; reverse-flush column with strong solvent to dissolve blockage
Column particle size too small (for example 3 micrometers) Use larger particle size (for example 5 micrometer)
Microbial growth on column  Use at least 10% organic modifier in mobile phase; use fresh buffer daily; add 0.02% sodium azide to aqueous mobile phase; store column in at least 25% organic solvent without buffer
Mobile phase viscosity too high  Use lower viscosity solvents or higher temperature 
Plugged frit in in-line filter or guard column Replace frit or guard column
Plugged inlet frit  Replace endfitting or frit assembly
Polymetric columns - solvent change causes swelling of packing  Use correct solvent with column; change to proper solvent compositionl consult manufacturer's solvent-compatibility chartl use a column with a higher percentage of cross-linking
Salt precipitation (especially in reversed-phase chromatography with high concentration of organic solvent in mobile phase)  Ensure mobile phase compatibility with buffer concentration; decrease ionic strength and water-organic solvent ratio; premix mobile phase
When injector disconnected from column - blockage in injector  Clean injector or replace rotor