Changing retention times: Causes and Solution

 

                           Possible Cause                                                                                            Solution

Buffer retention times Use buffer with concentration greater than 20 mM.
Contamination buildup Flush column occasionally with strong solvent
Equilibration time insufficient for gradient run or changes in isocratic mobile phase Pass at least 10 column volumes through the column for gradient regeneration or after solvent changes
First few injections - active sites Condition column by injecting concentrated sample
Inconsistent on-line mobile-phase mixing Ensure gradient system is delivering a constant composition; compare with manually prepared mobile phase; partially premix mobile phase
Selective evaporation of mobile-phase component Cover solvent reservoirs; use less-vigorous helium purging; prepare fresh mobile phase
Varying column temperature Thermostat or insulate column; ensure laboratory temperature is constant.